UL 61010A-2-041-2002《实验室用电气设备 第2部分 医学材料处理和实验室材料处理用高压蒸汽灭菌器的具体要求》
1.1 Scope
Replacement:
This standard applies to AUTOCLAVES,including those with an automatic loading and unloading system,
which incorporate a PRESSURE VESSEL using steam within the absolute pressure range from 0 to 500 kPa,
and intended for the treatment of medical materials and for laboratory processes,for example for
ster…zation
NOTES
1 National and other regulations or codes apply for the safety of automatic loading and MRloading systems.
2 All pressures are specified in absolute terms Atmospheric pressure(1 bar)一100 kPa
Where an AUTOCLAVE incorporates a steam generator having its own PRESSURE VESSEL within the same
enclosure,the applicable safety requirements for the AUTOCLAVE PRESSURE VESSEL apply also to the steam
generator(see 14 1 051
1 1 2 Equipment excludedfromthe scope
Addition:
Add the following new dash
pressu rized environmental cabinets
NOTES
1 Since all AUTOCLAVES covered by this part 2 use steam,other types using dry heat,toxic gas or radiation are not
included.