UL 61010A-2-042-2002《实验室用电气设备 第2部分 医学材料处理和实验室材料处理用使用有毒气体的高压灭菌器和杀菌器的具体要求》
1.1 Scope
Replacement:
This standard applies to AUTOCLAVES and STERILIZERS,including those with an automatic loading and
unloading system,which incorporates a CHAMBER using toxic gas intended for the treatment of medical
materials,and for laboratory processes,for example for sterilization
NOTES
1 National and other regulations or codes apply for the safety of automatic loading and MRloading systems
2 In general it is considered that automatic control of the sterilization cycle is essential for the safe operation of
equipment using toxic gases because a manual control system might present serious hazards(see 1 2)to the
0PERATOR
3 The principal sterilant gases in use are ethylene oxide and formaldehyde
4 For some applications,the CHAMBER may be operated at a pressure above or below atmospheric pressure.
5 All pressures are specified in absolute terms Atmospheric pressure(1 bar)≥100 kPa
6 The word AUTOCLAVES includes STERILIZERS unless specifically stated otherwise.
Where an AUTOCLAVE
标准信息:
- 免责声明:本网站所提供的电子文本仅供参考,请以正式标准出版物为准。本网站所提供的电子文本均来自于互联网,其版权由其发行商所有,仅供个人学习、研究之用,未经授权,禁止复制、发行、汇编、翻译等。
- 中文名称:实验室用电气设备 第2部分 医学材料处理和实验室材料处理用使用有毒气体的高压灭菌器和杀菌器的具体要求
- 英文名称:UL Standard for Safety Electrical Equipment for Laboratory Use; Part 2: Particular Requirements for
- 国际标准分类:11.080.10
- 中国标准分类:N61
- 标准前缀:UL
- 标准状态:废止
- 发布单位:US-UL
- 发布日期:无
- 废止日期:无
- 实施日期:无
- 页数:0 页